JOI Alliance (IODP-USIO) home JOI Alliance (IODP-USIO) employee intranet JOI Alliance (IODP-USIO) staff directory JOI Alliance (IODP-USIO) web site map Search the JOI Alliance (IODP-USIO) web sites
About the IODP-USIO
JOI Alliance (IODP-USIO): expedition and participant information
JOI Alliance (IODP-USIO) core/log databases and core sample curation
JOI Alliance (IODP-USIO) drilling/logging tools and science laboratories
JOI Alliance (IODP-USIO) publications
JOI Alliance (IODP-USIO) educational resources and outreach activities
JOI Alliance (IODP-USIO) news releases, photo gallery, and promotional material
JOI Alliance (IODp-USIO) meetings, port calls, and travel information
JOI Alliance (IODP-USIO) employment opportunities
JOI Alliance (IODP-USIO) contact information
Tools & labs > Logging tools

Downhole logging tools

Triple Combo Tools

Temperature/Acceleration/Pressure Tool (TAP)

Description

    The TAP (High Resolution Temperature / Acceleration / Pressure) tool was designed at LDEO to acquire borehole temperature, tool acceleration and hydrostatic pressure data. It is the successor to the Lamont Temperature Tool (TLT), which was used in the early part of ODP.

    The TAP tool is fastened to the bottom of the Triple Combo; because it is a memory tool, the stored data are dumped to the third party DAS (Data Acquisition System) upon the tool's return to the rig floor.

    Fast and slow response thermistors are mounted near the bottom of the tool to detect borehole fluid temperatures at two different rates. The thinner, fast-response is able to detect small abrupt changes in temperature, the thicker, slow-response thermistor is used to estimate temperature gradient and thermal regimes more accurately. One pressure transducer is included to turn the tool on and off at specified depths when used in memory mode. Typically data acquisition is programmed to begin 100 m above the sea floor.

    A 3-axis accelerometer is also included to measure tool movement downhole. These data are expected to be instrumental in analyzing the effects of heave on a deployed tool string, which will lead to the fine tuning of the WHC (wireline heave compensator).


Applications

    Geothermics
    The recording of temperature provides an insight into the thermal regime of the formation surrounding the borehole. The vertical heat flow is estimated from the vertical temperature gradient combined with the measurements of the thermal conductivity from logs or core samples.

    Hydrogeology
    Crust at mid-ocean ridge crests must be permeable to a considerable depth to allow for the efficient removal of heat by hydrothermal systems. Temperature logs in such an environment can clearly differentiate between the advective (hydrothermal) and conductive heat transfer regimes.


Environmental Effects

    Drilling and circulation operations considerably disturb the temperature distribution inside the borehole thus preventing equilibrated temperature conditions. The amount of time elapsed between the end of drilling fluid circulation and the beginning of logging operations is not long enough to allow the borehole to recover thermally. Therefore the data recorded is not representative of the thermal equilibrium of that environment. In addition, the thermistors may become fouled with sediment from the drilled formation which reduces the sensitivity and accuracy of the recorded temperature data.


Log Presentation

    Temperature data acquired by the fast and slow thermistors may be presented with resistivity, density and porosity log data, or simply alone. For an example of temperature profile, click here.


Tool Specifications

Temperature rating: 185° F (85° C)
Pressure rating : 10 kpsi (69 kPa)
Diameter: 3.25 in (8.26 cm)
Length: 8.895 ft (2.71 m)
Temperature sampling rate: 1 Hz
Acceleration sampling rate:
Low resolution mode (LR): 4 Hz
High resolution mode (HR): 8 Hz
Pressure sampling rate: 1 Hz
Total data recording time:
HR mode: 5 hrs.
LR mode: 8 hrs.
Power source: 8 alkaline batteries (D type)
Operation time from one set of batteries: approx. 40 hrs.


Measurement Specifications

Temperature measurement range: 25-185° F (-4 to 85° C)
Temperature resolution: 0.002° F (0.001° C)
Temperature accuracy: ± 3.6° F (± 2° C)
Acceleration measurement range:  -20 to 20 m²
Acceleration resolution: 1e-³ m/s²
Acceleration accuracy: ± 0.6 m/s²
Pressure measurement range: 0-10 kpsi (4.5 kPa)
Pressure resolution: 1 psi (0.45 Pa)
Pressure accuracy: ± 1.4 MPa (± 2% FS)


Deployment Notes

    Because it is a memory tool, the TAP needs to be initialized approximately 1/2 hour prior to rig up of the Triple Combo tool string. Once initialized, the TAP tool should be placed on the deck outside of the DHML to be picked up by the roughnecks. The logger then must connect it to the bottom of the Triple Combo using a pin and rotating ring assembly. When the Triple Combo is retrieved to the rig floor, the Logging Staff Scientist must remove the TAP tool, wash it off and download the data on the onboard computer.